If transmitter is unable to be tuned correctly, return the
transmitter to the ESC for repair.
8
Retest the transmitter using the Power Output Test. If the
transmitter fails, return the transmitter to the ESC for repair.
9
Perform field current drain test (refer to
Output Power Test and Center Frequency Tests For Models
91341-09, 91343-09, and 91347-09
Figure 4-15 9134x-09, parts needed for tests.