Cleaning/Disinfecting
To clean the exterior of monitors, modules, and cables:
•
Prepare the cleaning solution according to the manufacturer’s
instructions.
•
Wet a clean cloth with the selected cleaning solution.
•
Remove excess liquid from the cloth and squeeze dry.
•
Wipe exposed surfaces of the equipment and cables.
•
Remove any soap residue by gently wiping with a clean damp
cloth.
•
Wipe dry with a clean dry cloth.
Warnings:
•
Use only recommended cleaning solutions, or you may void the
manufacturer’s warranty.
•
Harsh chemical agents degrade plastics and will compromise
the safety of the device. Some germicidal and other harsh
cleaning compounds are known to damage some plastics by
weakening the structural integrity and compromising the
electrical insulating properties.
•
Disconnect the equipment from the patient and the electrical
supply before cleaning.
•
Do not allow liquid to enter the interior of the transmitter,
module, or other monitoring equipment.
•
Do not use any transmitter or equipment that has been
compromised by liquid ingress.
•
Do not immerse the equipment or cables in water or cleaning
solutions.
•
Do not autoclave.
•
Do not allow any liquid to enter the case. An open battery
compartment is not fluid resistant.
Cautions:
•
Use caution when cleaning cable connectors so that liquid is not
permitted to collect around the electrical contacts or seep
inside the connector. Trapped liquids and surface residues
provide an unintentional electrical path, which may cause noisy
signals and false alarms.
•
Clean the transmitter case with the battery door closed. Fluids,
including cleaning solutions, may damage the electrical
components inside and cause the transmitter to malfunction.
•
Observed effects of liquid ingress include delays in reporting
SpO
2
and NIBP measurements and incorrect channel frequency
programming on the 9134x 1400-MHz transmitters.