Multiparameter Transmission (90343/91343)
The following data can be transmitted by the multiparameter,
patient-worn transmitter:
•
SpO
2
(with Nellcor pulse oximetry sensors)
-
saturation
-
SpO
2
sensor status
-
pulse rate
•
NIBP (with the model 90217 ABP monitor)
-
systolic
-
diastolic
-
mean pressure
-
measurement time
-
alarm conditions
Battery Compartment SIP switches
SpO
2
data averaging is used to smooth the oximetry saturation
value by averaging the patient input values over 4, 8, or 16 seconds.
This selection is made by setting the SIP switches 1 and 2 beneath
the battery compartment in the multiparameter, patient-worn
transmitter. The default value is 8 seconds. Refer to
Notes:
•
Setting both SIP switches 1 and 2 to ON disables SpO
2
data
transmission.
•
To enable SpO
2
, remove the battery, set the selected interval,
and re-install the battery.
•
Disabling SpO
2
operation in the 90343/91343 patient-worn
transmitter lengthens battery life.
Caution:
Avoid using pencils or sharp instruments to configure the SIP
switches.
Table 1-4 SIP switch 1 and 2 Settings
SIP switch 1
SIP switch 2
Effect
OFF
OFF
4 seconds averaging enabled
OFF
ON
8 seconds averaging enabled (default)
ON
OFF
16 seconds averaging enabled
ON
ON
Disable SpO
2
operation