4 - 3 8
U
L T R A V I E W
D
I G I T A L
T
E L E M E T R Y
-
S
E R V I C E
M
A N U A L
M
A I N T E N A N C E
Initial Test Setup for 90341, 90347, 90343, 91341, 91343, and
91347 Transmitters
Figure 4-19 Test multimeter and the transmitters to be tested
1
Set the multimeter to milliamperes (mA), and then connect the
test leads to the current input jacks.
2
Insert the battery on its side, into the negative (-) side of the
battery holder. Connect the multimeter to the battery post (+)
using the positive lead. Connect the negative lead to the battery
contact of the transmitter being tested. Refer to
page 4-38 for multimeter probe positioning.
Figure 4-20 Probe connections