The following messages are displayed on the Ultraview monitor to
provide ABP status information to the caregiver. These messages
summarize the 90217 event codes. Some of these messages include
an event code in parentheses to provide more detailed analysis of
the event. A complete list of the event codes can be found in the
90217 ABP Monitor Operations Manual
(P/N 070-0137-xx), located
on CD-ROM P/N 084-0201-xx.
NIBP UNAVAILABLE (xx)
Displayed when the 90217 ABP monitor has detected an internal
condition, defined by the code (xx). Typically, this indicates a
hardware or software failure that requires that the transmitter be
removed from service.
NIBP READING FAILURE (xx)
Displayed when the ABP monitor was unable to make a reading. The
code (xx) defines the cause of failure.
NIBP AIR LEAK
Displayed when an air leak has been detected in the pneumatic
system, preventing a reading from being taken.
NIBP LOOSE OR NO CUFF
Displayed when the cuff was able to be inflated in a manner
indicating that it was not attached to the patient correctly.
NIBP READING CANCELLED
Displayed when the patient has pressed the START/STOP button on
the 90217 ABP monitor, halting a reading in progress.
NIBP LOW BATTERY
Displayed when the primary 90217 ABP monitor (3xAA) battery
voltage is low. Replace with fresh batteries.
NIBP KINKED HOSE
Displayed when the pressure value increased too rapidly indicating a
kinked hose or other restriction.
NIBP EVENT CODE (xx)
Displayed when the event code returned from the 90217 ABP
monitor is not defined into one of the other messages.