Figure 4-17 Side view of assembled parts with motor holder used as
stand-off
Output Test
To verify transmitter power ouput
1
Connect the assembled equipment to a spectrum analyzer.
2
Set up the spectrum analyzer as follows:
3
Verify that transmitter output power measures greater than -28
dBm.
CF
set to the frequency of the transmitter under test
Span
1 MHz/div
Ref Level
0 dBm
Video Filter
OFF
RBW
Auto
RF Attn
Auto