Maintenance
NIBP Calibration
Required tools and parts:
1.
Sphygmomanometer
2.
Pressure generator (e.g. squeeze bulb)
3.
Trimmer driver
4.
Two T-piece connector and 3 short hoses (approx. 15cm each)
5.
Adult cuff wrap on cylinder
6.
Stopwatch or equivalent (e.g. watch with second hand)
NIBP calibration
procedure
1.
Disassemble the monitor so that the NIBP calibration potentiometer R267 is accessible. This can
be done by removing the I/O Board from the rear enclosure and connect directly to the Main
Board.
2.
The position of R267 on the Main Board is shown below.
3.
Make sure the monitor can be powered on and can perform NIBP measurement.
4.
Setup the calibration system as show below.
4-33
NIBP calibration
potentiometer : R267
NIBP
reading on
display
Pressure of
Sphygmo-
manometer
Содержание elance 5
Страница 1: ...Vital Signs Monitor 93300 05 Service Manual 070 1949 00 Rev H...
Страница 7: ...Introduction External Recorder 1 4...
Страница 77: ...Glossary The following terms may appear in this manual 7 1...
Страница 78: ...Glossary 7 2...
Страница 79: ...Glossary 7 3...
Страница 85: ...Appendix B Symbols B 2...
Страница 86: ...Appendix B Symbols B 3...
Страница 87: ...Appendix B Symbols B 4...
Страница 88: ...Appendix B Symbols B 5...
Страница 89: ...Appendix B Symbols B 6...
Страница 90: ...Appendix B Symbols B 7...
Страница 91: ...Appendix B Symbols B 8...
Страница 92: ...Appendix B Symbols B 9...
Страница 94: ...Appendix B Symbols lance is a registered trademark of Spacelabs Healthcare B 11...