Theory
SpO
2
Block Diagram
-
Patient SpO
2
detection is by comparison of Red and IR light absorption by patient arterial blood.
-
The current for driving the LED is controlled by the ATMEGA162V microcontroller.
-
The signal detected at the SpO
2
sensor is amplified and captured at by the sample and hold
circuitry. Then, the signal is further amplified and sent through a 7-8Hz LFP.
-
Each of dynamic RED and IR and static RED and IR signals are compare with the ambient signal
at the Differential Amplified. This is to derive the SpO
2
IR, SpO
2
IR DC, SpO
2
RED, SpO
2
RED
DC signals.
-
The SpO
2
DC, SpO
2
IR, SpO
2
IR DC, SpO
2
RED, SpO
2
RED DC signals are sent to ATMEGA
microcontroller for processing.
3-5
Diff.
AMP
A
D
C
S/H
SpO
2
Input
S/H
S/H
LPF
LPF
Diff.
AMP
Diff.
AMP
Diff.
AMP
IR
DC
IR
Red
DC
Red
AMP
AMP
LPF
AMP
AMP
Red
IR
Ambient
Содержание elance 5
Страница 1: ...Vital Signs Monitor 93300 05 Service Manual 070 1949 00 Rev H...
Страница 7: ...Introduction External Recorder 1 4...
Страница 77: ...Glossary The following terms may appear in this manual 7 1...
Страница 78: ...Glossary 7 2...
Страница 79: ...Glossary 7 3...
Страница 85: ...Appendix B Symbols B 2...
Страница 86: ...Appendix B Symbols B 3...
Страница 87: ...Appendix B Symbols B 4...
Страница 88: ...Appendix B Symbols B 5...
Страница 89: ...Appendix B Symbols B 6...
Страница 90: ...Appendix B Symbols B 7...
Страница 91: ...Appendix B Symbols B 8...
Страница 92: ...Appendix B Symbols B 9...
Страница 94: ...Appendix B Symbols lance is a registered trademark of Spacelabs Healthcare B 11...