![Spacelabs Healthcare elance 5 Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/spacelabs-healthcare/elance-5/elance-5_service-manual_1339020011.webp)
Setup
Bedside Monitor Network Setup
Automatic IP Address Mode
1.
The monitor is default to DHCP mode and will be assigned an IP address automatically when
connected to a router or switch with DHCP server functionality.
2.
When the central notices a bedside monitor on the network, it displays a dialog box that allows the
operator to select the physical location of the bedside monitor.
Static IP Address Mode
1.
It is also possible to use static IP address for router and switch without DHCP server functionality.
To use the <Static IP address> mode, enter the biomed menu by touching <Monitor Setup> /
<Biomedical> / <Service> to access the <Setup Service> menu. Then select <Static IP> mode
from the <TCP/IP Setting>.
2.
The <Set Static IP> button will appear. Touch <Set Static IP> button to enter the <Set Static IP>
menu.
2-4
Содержание elance 5
Страница 1: ...Vital Signs Monitor 93300 05 Service Manual 070 1949 00 Rev H...
Страница 7: ...Introduction External Recorder 1 4...
Страница 77: ...Glossary The following terms may appear in this manual 7 1...
Страница 78: ...Glossary 7 2...
Страница 79: ...Glossary 7 3...
Страница 85: ...Appendix B Symbols B 2...
Страница 86: ...Appendix B Symbols B 3...
Страница 87: ...Appendix B Symbols B 4...
Страница 88: ...Appendix B Symbols B 5...
Страница 89: ...Appendix B Symbols B 6...
Страница 90: ...Appendix B Symbols B 7...
Страница 91: ...Appendix B Symbols B 8...
Страница 92: ...Appendix B Symbols B 9...
Страница 94: ...Appendix B Symbols lance is a registered trademark of Spacelabs Healthcare B 11...