Maintenance
Alarm Annunciation
•
Set ECG simulator to Asystole. Verify that the ECG numeric zone turn red and displayed
“ASY”, alarm is triggered with message “ECG asystole” displayed.
•
Set ECG simulator to VFib. Verify that the ECG numeric zone turn red and displayed “VF”,
alarm is triggered with message “ECG ventricular fibrillation” displayed.
•
Check: Life-threatening alarm tone is annunciated (in groups of ten at eight second intervals)
•
Record results
2nd Speaker
•
Check: 2nd speaker is activated if alarm is not attended in the preset period of time (0-3 minutes).
•
Check: Alarm Silence key (silences both speakers)
•
Set ECG simulator to normal sinus rhythm
•
Record results
Speaker Volume
•
Select “Speaker Volume” in <Monitor Setup> -> <Clinical> -> <System>
•
Check: speaker volume sounds at all 10 settings (note, only main speaker will sound)
Recorder Option
•
(Skip this page if Recorder not installed)
•
Connect Recorder
•
Check: élance reports ”Recorder connected”.
•
Open the recorder door
•
Check: élance reports “Recorder door is open”
•
Remove paper then close the recorder door
•
Check: élance reports “Recorder out of paper”
•
Insert a roll of paper and close door.
•
In élance <Monitor Setup> -> <Biomedical> -> <Service> Menu, run <Recorder Diagnostics>
•
Check: test pattern at end of recording for missing print head dots.
Power Cycle and Check Time
•
Power cycle monitor
•
Check: élance logo seen during start-up
•
Check: correct time on monitor display
Electrical Leakage Current Tests
•
Connect the monitor power cord and the patient leads to the safety analyzer.
•
Power up the Safety Analyzer
•
Power up the monitor
•
Test the monitor for normal and single fault conditions
•
Compare the results with Table 1 & 2 in the Electrical Safety Testing section above to insure safe
operation.
•
Power off the monitor
•
Equipment failing this test should be repaired before returning to service.
4-8
Содержание elance 5
Страница 1: ...Vital Signs Monitor 93300 05 Service Manual 070 1949 00 Rev H...
Страница 7: ...Introduction External Recorder 1 4...
Страница 77: ...Glossary The following terms may appear in this manual 7 1...
Страница 78: ...Glossary 7 2...
Страница 79: ...Glossary 7 3...
Страница 85: ...Appendix B Symbols B 2...
Страница 86: ...Appendix B Symbols B 3...
Страница 87: ...Appendix B Symbols B 4...
Страница 88: ...Appendix B Symbols B 5...
Страница 89: ...Appendix B Symbols B 6...
Страница 90: ...Appendix B Symbols B 7...
Страница 91: ...Appendix B Symbols B 8...
Страница 92: ...Appendix B Symbols B 9...
Страница 94: ...Appendix B Symbols lance is a registered trademark of Spacelabs Healthcare B 11...