![Spacelabs Healthcare elance 5 Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/spacelabs-healthcare/elance-5/elance-5_service-manual_1339020030.webp)
Maintenance
•
Select Respiration Lead in Channel 2
•
Disconnect LL (Red)
–
Observe that the RESP waveform disappears
•
Reconnect LL, Disconnect RA (White)
–
Observe that the RESP disappear
–
Reconnect RA
•
Record results
Temperature
•
Connect a temp probe to the monitor.
•
Leave the temp probe in room temp for 3 min.
•
Observe that the monitor displays the room temp.
•
Hold the temperature probe and wait for 3 min.
•
Observe that the monitor displays approx. 36.5 degrees Celsius or your body surface temperature.
IBP
•
Set P1 IBP Systolic alarm limit to 110(L) and 130(H), Diastolic alarm limit to 70(L) and 90(H),
Mean alarm limit to 85(L) 105(H).
•
Set the simulator to generate Systolic pressure below 110mmHg. Observe that the monitor
generates P1 IBP Systolic low limit alarm.
•
Set the simulator to generate Systolic pressure above 130mmHg. Observe that the monitor
generates P1 IBP Systolic high limit alarm.
•
Set the simulator to generate Diastolic pressure below 70mmHg. Observe that the monitor
generates P1 IBP Diastolic low limit alarm.
•
Set the simulator to generate Diastolic pressure above 90mmHg. Observe that the monitor
generates P1 IBP Diastolic high limit alarm.
•
Set the simulator to generate Mean pressure below 85mmHg. Observe that the monitor generates
P1 IBP Mean low limit alarm.
•
Set the simulator to generate Mean pressure above 105mmHg. Observe that the monitor
generates P1 IBP Mean high limit alarm.
•
Same setting and test apply to P2.
EtCO2
•
Connect the EtCO2 sampling line to a 5% CO2 gas cylinder output port. Verify that the EtCO2
reading on the monitor is 38mmHg
±
2mmHg.
•
Set the EtCO2 alarm high limit to 30mmHg. Verify that the monitor generates EtCO2 high limit
alarm.
•
Set the EtCO2 alarm low limit to 45mmHg. Verify that the monitor generates EtCO2 low limit
alarm.
Network
•
Check: Central Waveforms match monitor
•
Check: Central Parameters match monitor
•
Record results
4-7
Содержание elance 5
Страница 1: ...Vital Signs Monitor 93300 05 Service Manual 070 1949 00 Rev H...
Страница 7: ...Introduction External Recorder 1 4...
Страница 77: ...Glossary The following terms may appear in this manual 7 1...
Страница 78: ...Glossary 7 2...
Страница 79: ...Glossary 7 3...
Страница 85: ...Appendix B Symbols B 2...
Страница 86: ...Appendix B Symbols B 3...
Страница 87: ...Appendix B Symbols B 4...
Страница 88: ...Appendix B Symbols B 5...
Страница 89: ...Appendix B Symbols B 6...
Страница 90: ...Appendix B Symbols B 7...
Страница 91: ...Appendix B Symbols B 8...
Страница 92: ...Appendix B Symbols B 9...
Страница 94: ...Appendix B Symbols lance is a registered trademark of Spacelabs Healthcare B 11...