Maintenance
Replacing the Inverter board Assembly
1.
Disconnect battery power by moving the rear switch to off position.
2.
Remove the Front Panel.
3.
Remove the Main Board and Zbitsy Board from the rear enclosure.
4.
Remove the Mylar sheet covering the battery.
5.
Remove two screws that hold the Inverter Board on the rear enclosure.
Replacing the Power Supply Assembly
1. Disconnect battery power by moving the rear switch to off position.
2. Remove the Front Panel.
3. Remove the Main Board and Zbitsy Board from the rear enclosure.
4. Remove the Mylar sheet covering the battery.
5. Remove four screws that hold the Power Supply Board on the rear enclosure.
Remove Screw
4-21
Remove Screw
Remove Screw
Remove Screw
Remove Screw
Remove Screw
Содержание elance 5
Страница 1: ...Vital Signs Monitor 93300 05 Service Manual 070 1949 00 Rev H...
Страница 7: ...Introduction External Recorder 1 4...
Страница 77: ...Glossary The following terms may appear in this manual 7 1...
Страница 78: ...Glossary 7 2...
Страница 79: ...Glossary 7 3...
Страница 85: ...Appendix B Symbols B 2...
Страница 86: ...Appendix B Symbols B 3...
Страница 87: ...Appendix B Symbols B 4...
Страница 88: ...Appendix B Symbols B 5...
Страница 89: ...Appendix B Symbols B 6...
Страница 90: ...Appendix B Symbols B 7...
Страница 91: ...Appendix B Symbols B 8...
Страница 92: ...Appendix B Symbols B 9...
Страница 94: ...Appendix B Symbols lance is a registered trademark of Spacelabs Healthcare B 11...