Maintenance
Replacing the CO
2
Board
1.
Disconnect battery power by moving the rear switch to off position.
2.
Remove the Front Panel.
3.
Remove the Main Board and Zbitsy Board from the rear enclosure.
4.
Remove four screws that hold the CO
2
Board on the rear enclosure.
5.
Replace EtCO
2
Board and reassemble the monitor.
4-23
Remove Screw
Remove Screw
Remove Screw
Remove Screw