Maintenance
Assembly/Disassembly Procedures
Caution:
Before beginning any disassembly procedures, power OFF the monitor,
disconnect the AC cord from the AC power receptacle, remove the battery switch
label and disconnect the battery by moving the battery switch to
position.
Observe precautions for handling electrostatic-sensitive devices!
Replace the battery switch label when the monitor is re-assembled.
Proper torque 6.0
±
±
±
±
0.5 lb-in is recommended to secure the enclosure with screws at
assembly.
Required Tools and Parts
• Anti-static mat with wrist strap
• #1 and #2 Phillips-head screwdriver
• 5/16-inch nut driver
• 3/16-inch nut driver
Installing or Replacing the Optional Recorder
1.
Align the top and bottom locking pins with the keyholes on the monitor side panel.
2.
Side the recorder downward until the latch locks the recorder in position.
3.
To uninstall, release the latch and slide the recorder upward until it reach the top of the keyhole.
Then, gently detach it from the monitor side panel.
4-9
Latch
Latch Release
Содержание elance 5
Страница 1: ...Vital Signs Monitor 93300 05 Service Manual 070 1949 00 Rev H...
Страница 7: ...Introduction External Recorder 1 4...
Страница 77: ...Glossary The following terms may appear in this manual 7 1...
Страница 78: ...Glossary 7 2...
Страница 79: ...Glossary 7 3...
Страница 85: ...Appendix B Symbols B 2...
Страница 86: ...Appendix B Symbols B 3...
Страница 87: ...Appendix B Symbols B 4...
Страница 88: ...Appendix B Symbols B 5...
Страница 89: ...Appendix B Symbols B 6...
Страница 90: ...Appendix B Symbols B 7...
Страница 91: ...Appendix B Symbols B 8...
Страница 92: ...Appendix B Symbols B 9...
Страница 94: ...Appendix B Symbols lance is a registered trademark of Spacelabs Healthcare B 11...