Maintenance
IBP Calibration
Required tools and parts:
1.
Patient Simulator (Dynatech Nevada 300B or equivalent) with Invasion BP cable, or
2.
Reusable pressure transducer, or
3.
Disposable pressure transducer
IBP calibration
procedure
1.
Perform IBP calibrations only on units with option ‘I’ whenever the Zbitsy Board or Main Board
is replaced.
2.
Plug the two pressure cables (BP1/BP2) from the patient simulator into the IBP1/IBP2 connectors
of the Patient Input Panel of the monitor. Set the simulator to a static pressure of 0mmHg for
both BP1 and BP2. Or connect pressure transducer to the IBP1/IBP2 connectors.
3.
Power on the monitor.
4.
Select <MONITOR SETUP> / <Biomedical> / <Service> keys.
5.
Select <Start P1 Calibration>. The message “P1 calibrating” will be displayed at the bottom of the
screen. Waiting briefly. Verify message “P1 calibration OK” followed by 2 beep tones.
6.
Select <Start P2 Calibration>. The message “P2 calibrating” will be displayed at the bottom of the
screen. Waiting briefly. Verify message “P2 calibration OK” followed by 2 beep tones.
4-32
Содержание elance 5
Страница 1: ...Vital Signs Monitor 93300 05 Service Manual 070 1949 00 Rev H...
Страница 7: ...Introduction External Recorder 1 4...
Страница 77: ...Glossary The following terms may appear in this manual 7 1...
Страница 78: ...Glossary 7 2...
Страница 79: ...Glossary 7 3...
Страница 85: ...Appendix B Symbols B 2...
Страница 86: ...Appendix B Symbols B 3...
Страница 87: ...Appendix B Symbols B 4...
Страница 88: ...Appendix B Symbols B 5...
Страница 89: ...Appendix B Symbols B 6...
Страница 90: ...Appendix B Symbols B 7...
Страница 91: ...Appendix B Symbols B 8...
Страница 92: ...Appendix B Symbols B 9...
Страница 94: ...Appendix B Symbols lance is a registered trademark of Spacelabs Healthcare B 11...