1.5
Location Installation Considerations
The Southland Sensing OMD-525X Oxygen Analyzer is designed to be wall mounted in a general
purpose area and uses a NEMA4X / IP66 Enclosure. When installed outdoors in cold areas an optional
heater is recommended as well. Consider also giving the analyzer a sun shield if it is going to be
mounted in the direct sunlight.
Reference your local electrical authority for the proper installation.
The analyzer is EMI / RFI protected, however it is good practice not to mount it too close to sources
of electrical interference such as large transformers, motor start contactors, relays, large pumps, etc.
Also avoid subjecting the analyzer to significant vibration.
Mount the unit at a suitable eye level to easily read the local display. Gas connections are located on
the top right side of the analyzer, ensure there is adequate room to hook up your gas lines.
1.6
Safety Considerations
The oxygen analyzer is designed for installation into a general purpose area. Please follow your local
electrical code.
Designed, Tested, and Assembled in California, USA
4045 E. Guasti Rd. #203 Ontario, CA 91761 USA : 1-949-398-2879 : [email protected] : www.sso2.com
Introduction
OMD-525X Oxygen Analyzer