Note: We have tested 100s of USB flash drives and 99% of them work just fine. Prior to data logging
important data, make sure to test your USB thumb drive and verify it is logging. The USB Icon should
appear or disappear as the USB stick is installed or removed. If for some reason the electronics don’t
recognize the USB stick, consider restarting the analyzer to see if the USB Icon appears.
To install a USB flash drive, simply plug it into the left side of the display panel. This port will allow
the electronics to talk to your USB flash drive and save the stored data in an .csv file (excel compat
-
ible).
The data is stored in files by date and can be configured to save data at 1 minute intervals up to 120
minutes.
To select your data logging interval, simply go into the MENU screen and select LOGGING INTERVAL.
this will prompt you to a screen where you can set the value by minutes. Once selected, press
enter and your interval will be saved.
For data logging purposes, it is important to correctly set the date and time so that the time stamp
will be saved along with the recorded oxygen value. To set the time and date, refer to section 3.2.
3.9
USB Data Logging
Operation
OMD-525X Oxygen Analyzer
Designed, Tested, and Assembled in California, USA