** Caution: RANGE ID converters are internally powered and do not require external power.
DO NOT supply any voltage across these terminals as the output will be damaged. It is also
important to assure proper grounding of the external recording device.
Refer to section 2.3 for electrical wiring diagram.
When connecting the RANGE ID, connect to the (+) positive and (-) negative sides of the RANGE
output on the rear of the analyzer. To verify the signal output of the RANGE ID is working properly,
connect a voltmeter across the terminals to verify. You can change the range manually in MANUAL
RANGE mode to verify:
PPM Config.
Percent Config.
Output
Range 1
0 - 10 ppm
0 - 1%
1 VDC
Range 2
0 - 100 ppm
0 - 5%
2 VDC
Range 3
0 - 1000 ppm
0 - 10%
3 VDC
Range 4
0 - 10000 ppm
0 - 25%
4 VDC
Range 5
0 - 25%
0 - 100%
5 VDC
3.6
Range ID
Operation
OMD-525X Oxygen Analyzer