![Southland OMD-525X Скачать руководство пользователя страница 3](http://html1.mh-extra.com/html/southland/omd-525x/omd-525x_instruction-manual_1337452003.webp)
1.1
General Introduction
The Southland Sensing OMD-525X Oxygen Analyzer is a microprocessor based online unit designed for
continous measurements in a variety of applications and gas mixtures.
The analyzer was designed with the customer in mind, keeping the operations simple while still featur-
ing a fast response and rugged design. Every effort has been made to use modern industrial compo
-
nents and materials which has resulted in an advanced design, excellent performance, and an overall
low cost of ownership.
The analyzer has numerous options including PPM, Percent, and Purity oxygen analysis. The unit can
be configured for VAC or VDC power input with an analog 4 - 20mA concentration output and digital
RS485 Modbus ASCII output meeting most general purpose oxygen analysis applications.
Southland Sensing Ltd. appreciates your business and recommends reading through the complete
manual to be able to get the full experience from your new oxygen analyzer.
1.2
Principle of Operation - The Oxygen Sensor
The precision electrochemical oxygen sensor used in the OMD-525X is designed and manufactured by
Southland Sensing Ltd. under a strict quality procedure.
To understand how the oxygen analyzer functions, it is important to understand a little bit of the sen-
sor characteristics.
The active components in the precision electrochemical oxygen sensor are the anode, cathode, and
aqueous electrolyte which are all housed in the cell body. The oxygen molecules in the application
pass through the front sensing membrane into the electrolyte, where a chemical reaction occurs and a
raw electrical current is generated.
This electrical current is proportional to the amount of oxygen in the application. The analyzer then
processes this raw electronic signal, compensates for temperature and barometric pressure variations,
and converts the data into a parts-per-million or percent oxygen measurement value.
The oxygen concentration reading is then displayed in real time on the full backlit display and the user
can automate the control of their process using the standard 4 - 20 mA concentration analog output
signal which can be run to a PLC or other type of DCS system.
Precision Electrochemical Oxygen Sensor
Introduction
OMD-525X Oxygen Analyzer
Designed, Tested, and Assembled in California, USA
4045 E. Guasti Rd. #203 Ontario, CA 91761 USA : 1-949-398-2879 : [email protected] : www.sso2.com