Operation
How to use the FUNCTION field
Thermo Scientific
WX Ultra Series
2-57
Step
Key operation
Screen display and considerations
1
Set the Speed in the Run Screen.
>> See Section
Setting run conditions
.
2
(1) Use cursor keys to move the cursor
to RCF • w
2
T
on the Run Screen and
press the ENTER key.
(2) Use cursor keys to move the cursor
to w
2
T
and press the ENTER key.
3
Use cursor keys to move the cursor to
the w
2
T
setting field and enter a
desired w
2
T
value.
Example: 9.87 x 1011
4
When you press START key, the
machine starts to perform an w
2
T
run.
• The contents of the display
in the FUNCTION field
change.
•
ω
2
T
blinks.
• The FUNCTION field
switches to the
ω
2
T
field.
ω
2
T
display field
ω
2
T
setting field
• The value you entered is
displayed in the
ω
2
T
setting
field.
• The time value is calculated
from the speed and the
ω
2
T
value and the run time is
displayed in the Time Setting
field.
• Arithmetic and integral
operations are performed in
the
ω
2
T
display field.
ω
2
T display field