Operation
Print utilities (option)
Thermo Scientific
WX Ultra Series
2-87
1) Automatic printing of run records
Ultracentrifuge run records are retained in the run record buffer. This buffer can hold a maximum of
40 records at any given time. When the buffer becomes full, all of the 40 records are automatically
printed out, except when the PRINTOUT AFTER EACH RUN options is toggled on. In this case,
the result of the last run made is printed out, followed by all the run records in the buffer.
2) Automatic printing of run records of rotor
When the number of runs using a particular rotor with the ultracentrifuge has reached 20, the retained
run records for the rotor will be automatically printed out.
3) Printout after each run
When the PRINTOUT AFTER EACH RUN option is toggled on, the result of a run is automatically
printed out upon completion of that run. The printout produced is exactly the same as that produced
by selecting “5. RUN RESULT PRINTOUT”.
Printer fault handling
If a fault condition occurs that causes the printing in progress to be interrupted, do the following:
Note
When the RECORD AFTER EACH RUN option is toggled off, ultracentrifuge runs are not
recorded in the run record buffer (See
page 2-79
). Therefore, even if the 40th run in a sequence is
completed, the run records, including the 40th, will not be automatically printed out.
Contents of error
Process
(Executing the following processing permits printing out from the beginning.)
1
Centrifuge power OFF
The record is left in the centrifuge memory. After power is restored, you can print
them.
2
Printer power OFF
(Dead battery)
The power lamp blinks at intervals of about 0.5 sec and the printer is put into an
offline mode. Connect the AC adapter attached to the printer.
When data remains in the buffer memory, the online lamp blinks. Do not turn off the
power supply but connect the AC adapter as quickly as possible. In the online, the
remaining data will be printed out.
3
No print paper
The offline lamp blinks.
When paper is set, the offline lamp changes from the blinking state to the lighting
state.
Press the online switch, and the printer will start to print.
Note
By clearing the “printer paper exhausted” fault, you can continue the operation of the
ultracentrifuge. However, if the number of runs of the ultracentrifuge has reached 40, or if the
number of runs of the rotor in use has reached 20, then continuing the operation will result in the
loss of the oldest record.