Operation
How to use the FUNCTION field
Thermo Scientific
WX Ultra Series
2-43
4
Use cursor keys and numeric keys to
enter month, day, hours, and minutes.
Then, press ENTER key.
>> To set the hours, use a number between 0 and 23 (in the 24-hour system).
>> Enter a time later than the current time.
When setting the “Setting stop time”, allow for centrifugation time and set the
centrifugation start time to a time later than the current time.
>> You cannot set the centrifugation start time to a date more than 20 days
ahead.
5
If you are sure that you have made the
correct settings, select Yes.
Yes
No
>> If you select “Yes”
>> If you select “No”, you will get the display you see in process 2.
Make the selection again.
6
Check the RTC setting in the Run
Screen, then press START key.
The RTC run does not get activated
unless you press START key.
>> Run this machine according to Section
Normal operation
.
>> Once the RTC setting is made, you cannot change the run time (centrifugation
time).
If you wish to make a change to the run time, cancel RTC first.
>> Pressing START enters the system into a “DELAY” (and vacuum comes on),
condition causing the centrifuge to wait until the set time comes. This centrifuge
will turn the rotor automatically at the specified time and run it for the specified
period.
• The display switches to the
Run Screen and the
specified RTC time
appears.
• The message field displays
RTC.
• The run mode in the message
display field switches to
“DELAY”.
• Blink the PROG • RTC
in the FUNCTION field.
• To specify the run time
(centrifugation time), do
not set HOLD. Use ten
keys to enter the run time
instead.