Operation
How to use the FUNCTION field
Thermo Scientific
WX Ultra Series
2-53
How to set an RCF value
Enter a rotor number and an RCF value, and the machine will calculate, set, and display the speed.
Shown below is the procedure.
Step
Key operation
Screen display and considerations
1
While in the Run Screen, set a speed.
>> See Section
Setting run conditions
.
2
(1) While in the Run Screen, use cursor
keys to move the cursor to RCF •
ω
2
T
and press ENTER key.
(2) Move the cursor to RCF
max
(or
RCF
avg
) and press ENTER key.
3
Check that ROTOR LIST blinks and
press the ENTER key.
• The FUNCTION field
switches to what is shown
in the diagram to the left.
• The RCF
max
field blinks.
• The FUNCTION field
switches to the RCF field.
• The ROTOR LIST field
blinks.
• The display switches to the
Choose Rotor screen.
Содержание WX 100 Ultra