Operating Manual | 8 Troubleshooting
SONASCREEN
84 / 89
Revision: 1.0 | 2022-03-04
8
Troubleshooting
Potential error
Possible cause
Troubleshooting
After full reset, the "Leakage
measurement“ icon is not
shown on the "Start" screen.
After full reset, all
parameters and connections
of the device software have
been reset.
The information that leakage
measurements are possible
with the device is stored in
the device's hardware.
After full reset, the device software
must connect again to the device
hardware to receive and store all
information required for leakage
measurements.
1. Open the "Measurement" screen.
2. Start a measurement.
3. End the measurement.
4. Switch to the "Start" screen.