Operating Manual | 4 Preparing a work order
SONASCREEN
37 / 89
Revision: 1.0 | 2022-03-04
4.3
Setting the trigger
Description
Triggers are used for automated saving of measurement sections during measurement
recording. Triggers for saving may be customized. This is possible with two different
methods.
Availability
Automated recording of measurement data with triggers is only available in the
"Measurement recording" mode.
Trigger method "BeamPic"
In the "BeamPic" trigger method, the trigger is determined within the beam picture. The
following parameters may be set:
Parameter
Description/function
Pre-trigger time
Duration in seconds before triggering a recording. This duration is part
of the section of the current measurement that is to be saved.
Post-trigger time
Duration in seconds after triggering a recording. This duration is part
of the section of the current measurement that is to be saved.
Data to store
Determination of data that are to be saved automatically:
•
Measurement:
the measurement data of the set recording time
will be saved
•
Screenshot:
a screenshot of the current beam picture will be
created and saved
•
Measu Screenshot:
the measurement data of the
measurement duration and a screenshot of the current beam
picture will be saved
Trigger level
Limit value for triggering the recording, in dB.
The limit value must be within a range between upper and lower cutoff
frequency.
The recording starts as soon as the defined trigger level is reached or exceeded during the
current measurement. A marker is automatically set at the start of the recording.
After the defined post-trigger time, a section of the current measurement will automatically
be saved. This section contains all data to be saved within the defined pre- and post-trigger
time.
The appropriate trigger level may for instance be determined in a previous
measurement.
Setting the "BeamPic" trigger method
1.
Open the "Configuration" screen.