35. CHECKS AND ADJUSTMENTS
16. In this state, the following offset
values (tilt zero point error) are
calculated.
Xoffset = (X3+X4)/2
Yoffset = (Y3+Y4)/2
When both offset values fall within
the range ±20", adjustment is
completed.
Press
{ESC}
to return to <Instr.
const>.
If one of the offset values (Xoffset,
Yoffset) exceeds ±20", repeat the
check and adjustment procedures
from the beginning. If the
difference continues to exceed
±20" after repeating the check 2 or
3 times, have your local dealer
perform the adjustment.
With this option you can measure collimation error in your instrument so that the
CX can correct subsequent single face observations. To measure the error, make
angular observations using both faces.
PROCEDURE
1. Display <Collimation>.
Select “Instr.const” in the Config.
mode screen and select
“Collimation”.
2. Sight the reference point in Face
1.
Sight the reference point in Face 1
and press
[OK]
.
35.3 Collimation