Alarms & Troubleshooting
Battery Depleted
Two-tone alarm
Batteries are depleted and cannot operate pump; install
new batteries immediately.
Battery Removed Pump Won’t Run
Two-tone alarm
With AC adapter attached, batteries have been removed
while pump was running, or you have tried to start pump
with depleted batteries. Install new batteries
High Pressure
Two-tone alarm
High pressure is being caused by an occlusion in the
fluid path between the pump and the patient. Remove
the obstruction to resume operation. Or press NEXT or
STOP/START to stop the pump and silence the alarm for
two minutes, then remove occlusion and restart pump.
Upstream Occlusion
Two-tone alarm
Fluid is not flowing from the reservoir to the pump.
Check for a kink in the tubing between the reservoir and
pump. Remove the occlusion to resume operation. Press
NEXT or STOP/START to stop the pump and silence
alarm for two minutes, then remove the occlusion and
restart the pump.
Screen Displays Current Pump Status
Two-beep (long-short) alarm
The cassette is not aligned with the pump, or is damaged,
or a malfunction of the pump sensor(s) is occurring.
Reposition the pump to silence the alarm. If repositioning
the pump does not silence the alarm within two minutes,
the pump will display “No Disposable, Clamp Tubing.”
Air In Line Detected
Two-tone alarm
Air in fluid path. Press STOP/START or NEXT to silence
alarm. Make sure tubing is threaded in air detector
properly. If the fluid path contains air, close clamps,
disconnect from patient and prime fluid path.
Stopped
Three beeps every five minutes
Pump is stopped. Start pump, if appropriate.
Remote Dose Cord Removed
Two beeps when pump stopped, two-tone alarm when
pump is running
Remote dose cord removed while pump was running.
Programming Incomplete
Two-tone alarm when starting pump
Verify all programming screens by pressing ENTER/
CLEAR before moving to next screen or starting pump.
Read the entire Operator’s Manual before operating the CADD-Legacy
®
ambulatory
infusion pump. Failure to properly follow warnings, cautions and instructions could
result in death or serious injury to the patient. Please see the instructions for use for
a complete listing of the indications contraindications, warnings and precautions.
Warning: This Quick Reference Card should be used by clinicians only. Do not permit
patients to have access to this card, as the information would allow access to all
programming and operating functions. Assistance with the CADD-Legacy
®
PCA pump
is available to clinicians by calling 1-800-258-5361.