Change the Lock Level

1. 

Stop the pump.

2. 

Press LOCK.

3. 

Press 

or

 until desired lock level appears.

4. 

Press LOCK or ENTER/CLEAR.

5. 

Press 

or

 until the lock level code appears.

6. 

Press LOCK or ENTER/CLEAR.

Start the Pump

1. 

Press and hold STOP/START until 

(– – – – – – – – – – – –) disappears from the display.

2. 

Release STOP/START key. RUN will appear on the 

display when the pump is running.

Stop the Pump

1. 

Press and hold STOP/START until 

(– – – – – – – – – – – –) appears on the display.

2. 

Release STOP/START key. STOPPED will appear on 

the display when the pump is stopped.

Prime the Fluid Path

1. 

Pump must be stopped and in LL0 or LL1.  

Warning: Do not prime the fluid path with the tubing 

connected to a patient as this could result in over 

delivery of medication or air embolism.

2. 

Press and hold PRIME until the word PRIME appears 

on the display, along with (– – – – – – – – – –).

3. 

Release the PRIME key.

4. 

Press and hold PRIME to fill fluid path and eliminate 

air. Continue priming until the fluid path is free of air.

5. 

Press NEXT to return to the main screen.

Reset the Reservoir Volume

1. 

Stop the pump.

2. 

Press NEXT to display the Reservoir Volume screen.

3. 

Press ENTER/CLEAR to reset the value to previously 

programmed amount.

Turn the Pump On

1. 

Press and hold ON/OFF until pump beeps and 

powers up.

Turn the Pump Off

1. 

Press and hold ON/OFF until (•••• •••• ••••) 

appears on the display.

2. 

Release ON/OFF key.

3. 

The screen will go blank as the pump goes into a 

lower power state 

Change the Batteries

Pump Operations

IN19882EN-112017.indd   4

11/01/18   11:49

Содержание 6300

Страница 1: ...wer in use E Air Detector F Cassette part of reservoir or administration set that attaches to pump G Threaded Mounting Hole for use with Polemount Bracket Adapter back of pump H Battery Compartment back of pump I Cassette Latch attaches cassette to pump side of pump CADD Legacy PCA Infusion Pump Model 6300 Quick Reference Card for Clinicians Lock Level 0 LL0 G H I A B C E D F ...

Страница 2: ...ER CLEAR 9 Press NEXT Enter Continuous Rate 10 Press or to select desired Continuous Rate select the upper limit if the program will be adjusted in LL1 11 Press ENTER CLEAR 12 Press NEXT Warning This pump delivers medication at a constant rate and or allows delivery of a bolus dose at a specified time interval Programming the pump at a delivery rate other than what is prescribed will cause over or...

Страница 3: ...y will show 0 00 27 Press NEXT Verify Air Detector Status 28 Verify the setting is correct To change setting see Biomed Functions section 29 Press NEXT These screens will appear only if you have programmed a demand dose Verify Upstream Sensor Status 30 Verify the setting is correct To change setting see Biomed Functions section 31 Press NEXT Verify Programming 32 Press NEXT repeatedly to review pr...

Страница 4: ...path with the tubing connected to a patient as this could result in over delivery of medication or air embolism 2 Press and hold PRIME until the word PRIME appears on the display along with 3 Release the PRIME key 4 Press and hold PRIME to fill fluid path and eliminate air Continue priming until the fluid path is free of air 5 Press NEXT to return to the main screen Reset the Reservoir Volume 1 St...

Страница 5: ...Press or until the biomed code appears 4 Press LOCK or ENTER CLEAR 5 Press NEXT to go through settings 6 Press NEXT until AIR DETECTOR appears To Turn Air Detector ON HIGH ON LOW or OFF 7 Press or until desired setting appears 8 Press ENTER CLEAR 9 Press NEXT Warning When the air detector is turned off the pump will not detect air in the fluid path Periodically inspect the fluid path and remove an...

Страница 6: ...ir Detector is in use make a small loop of tubing underneath the air detector and hold it with your thumb Place the tubing over the groove in the air detector and tuck it under the catch Seat the tubing into the groove by gently pulling the tube upward until it is under the retention nubs and flat in the groove Warning Attach the cassette the part of the Medication Cassette reservoir or CADD admin...

Страница 7: ...emand doses given since screen was last cleared appears if demand dose is programmed Doses Attempted Displays number of demand doses attempted since screen was last cleared appears if demand dose is programmed Given Total amount of medication delivered since display was last cleared ml mg mcg Air Detector status only Indicates if air detector is turned on or off and high or low sensitivity level U...

Страница 8: ... demand dose press and hold STOP START to stop the pump To Start Clinician Bolus Pump must be running 2 Press LOCK 3 Press until Clinician Bolus Code appears 4 Press LOCK or ENTER CLEAR 5 Press or to select desired amount 6 Press DOSE or ENTER CLEAR to start bolus Note If you wish to stop delivery of a clinician bolus press and hold STOP START to stop the pump To Use the Remote Dose Cord Button 1 ...

Страница 9: ...to stop the pump and silence alarm for two minutes then remove the occlusion and restart the pump Screen Displays Current Pump Status Two beep long short alarm The cassette is not aligned with the pump or is damaged or a malfunction of the pump sensor s is occurring Reposition the pump to silence the alarm If repositioning the pump does not silence the alarm within two minutes the pump will displa...

Страница 10: ...r the batteries were removed within 15 seconds after stopping the pump Batteries must be reinstalled Important Always stop pump before removing batteries No Disposable Clamp Tubing Two beep long short alarm The disposable was removed or the cassette is not aligned with the pump or is damaged or a malfunction of the pump sensor s is occurring Clamp the tubing immediately Press STOP START or NEXT to...

Отзывы: