Change the Lock Level
1.
Stop the pump.
2.
Press LOCK.
3.
Press
or
until desired lock level appears.
4.
Press LOCK or ENTER/CLEAR.
5.
Press
or
until the lock level code appears.
6.
Press LOCK or ENTER/CLEAR.
Start the Pump
1.
Press and hold STOP/START until
(– – – – – – – – – – – –) disappears from the display.
2.
Release STOP/START key. RUN will appear on the
display when the pump is running.
Stop the Pump
1.
Press and hold STOP/START until
(– – – – – – – – – – – –) appears on the display.
2.
Release STOP/START key. STOPPED will appear on
the display when the pump is stopped.
Prime the Fluid Path
1.
Pump must be stopped and in LL0 or LL1.
Warning: Do not prime the fluid path with the tubing
connected to a patient as this could result in over
delivery of medication or air embolism.
2.
Press and hold PRIME until the word PRIME appears
on the display, along with (– – – – – – – – – –).
3.
Release the PRIME key.
4.
Press and hold PRIME to fill fluid path and eliminate
air. Continue priming until the fluid path is free of air.
5.
Press NEXT to return to the main screen.
Reset the Reservoir Volume
1.
Stop the pump.
2.
Press NEXT to display the Reservoir Volume screen.
3.
Press ENTER/CLEAR to reset the value to previously
programmed amount.
Turn the Pump On
1.
Press and hold ON/OFF until pump beeps and
powers up.
Turn the Pump Off
1.
Press and hold ON/OFF until (•••• •••• ••••)
appears on the display.
2.
Release ON/OFF key.
3.
The screen will go blank as the pump goes into a
lower power state