Display shows
programming information
and messages
B.
Power Jack used to plug in
AC Adapter
C.
Accessory Jack used to
plug in Remote Dose Cord
D.
Indicator Light indicates
AC power in use
E.
Air Detector
F.
Cassette (part of reservoir
or administration set that
attaches to pump)
CADD-Legacy
®
PCA
Infusion Pump Model 6300
Quick Reference Card for Clinicians Lock Level 0 (LL0)
For more information, please visit
www.smiths-medical.com
MPAUCA-1591
Alarms & Troubleshooting
Value Not Saved
A value was not saved by pressing ENTER/CLEAR. Press
NEXT to resume programming.
Error
Two-tone alarm
An error has occurred. Remove pump from service.
No Message Displayed
Two-tone alarm
The pump was running when batteries were removed or the
batteries were removed within 15 seconds after stopping
the pump. Batteries must be reinstalled.
Important: Always stop pump before removing batteries.
No Disposable, Clamp Tubing
Two beep (long-short) alarm
The disposable was removed, or the cassette is not aligned
with the pump or is damaged, or a malfunction of the pump
sensor(s) is occurring. Clamp the tubing immediately.
Press STOP/START or NEXT to silence the alarm.
Service Due
Two-tone alarm
The pump is due for service.
C
Read the entire Operator’s Manual before operating the CADD-Legacy
®
ambulatory
infusion pump. Failure to properly follow warnings, cautions and instructions could
result in death or serious injury to the patient. Please see the instructions for use for
a complete listing of the indications contraindications, warnings and precautions.
Warning: This Quick Reference Card should be used by clinicians only. Do not permit
patients to have access to this card, as the information would allow access to all
programming and operating functions. Assistance with the CADD-Legacy
®
PCA pump
is available to clinicians by calling 1-800-258-5361.