1.
Stop the pump.
2.
Push down and hold the arrow button on the battery
door while sliding the door off. Remove and discard
old batteries.
Important: Always stop pump before removing batteries.
3.
Install new batteries, matching polarities shown on
the pump. Replace battery door and close.
4.
Start the pump.
Warning: If a gap is present anywhere between the
battery door and the pump housing, the door is not
properly latched. If the battery door becomes detached
or loose, the batteries will not be properly secured
which could result in loss of power or non-delivery of
medication. Do not use rechargeable NiCad or nickel
metal hydride (NiMH) batteries. Do not use carbon zinc
(“heavy duty”) batteries.
Biomed Functions To Access
Pump must be stopped and in LL0.
1.
Press LOCK.
2.
Press LOCK or ENTER/CLEAR.
3.
Press
or
until the biomed code appears.
4.
Press LOCK or ENTER/CLEAR.
5.
Press NEXT to go through settings.
6.
Press NEXT until AIR DETECTOR appears.
To Turn Air Detector ON-HIGH, ON-LOW or OFF
7.
Press
or
until desired setting appears.
8.
Press ENTER/CLEAR.
9.
Press NEXT.
Warning: When the air detector is turned off, the pump
will not detect air in the fluid path.
Periodically inspect the fluid path and remove any air to
prevent air embolism.
To Turn Upstream Sensor ON/OFF
10.
Press NEXT until UPSTREAM SENSOR appears.
11.
Press
or
until desired setting appears.
12.
Press ENTER/CLEAR.
13.
Press NEXT.
Warning: When the upstream occlusion sensor is turned
off, the pump will not detect occlusions upstream
(between pump and reservoir). Periodically inspect the
reservoir for decreasing reservoir volume and inspect the
fluid path for kinks, a closed clamp, or other upstream
obstructions. Upstream occlusions could result in
under- or non-delivery of medications.
To Exit Biomed Functions
Pump Operations - Continued