The CADD-Legacy
®
PCA pump signals with audible
alarms and messages that appear on the display when a
condition requires your attention.
Res Vol Low
Three beeps
Level of fluid in the reservoir is low. Prepare to install
new reservoir.
Reservoir Volume Empty
Two-tone alarm
Reservoir volume has reached 0.0 ml. Press NEXT or
STOP/START to silence alarm. Install a new reservoir, if
appropriate.
Low Bat
3 two-tone beeps every five minutes
Batteries are low, but pump is operable. Change
batteries soon.
To Start Demand Dose
Pump must be running.
1.
Press DOSE or remote dose cord button.
Note: If you wish to stop delivery of the demand dose,
press and hold STOP/START to stop the pump.
To Start Clinician Bolus
Pump must be running.
2.
Press LOCK.
3.
Press
until Clinician Bolus Code appears.
4.
Press LOCK or ENTER/CLEAR.
5.
Press
or
to select desired amount.
6.
Press DOSE or ENTER/CLEAR to start bolus.
Note: If you wish to stop delivery of a clinician bolus,
press and hold STOP/START to stop the pump.
To Use the Remote Dose Cord/Button
1.
Use remote dose button on cord in same manner as
DOSE key on pump keypad.
2.
Attach the remote dose cord by inserting the
connector into the accessory jack and pushing firmly
until it snaps into place.
3.
Detach the remote dose cord by grasping the
connector and pulling it out of the jack.
Do not use excessive force or instruments, such as pliers,
to remove the remote dose cord from pump.