Batteries must be installed; pump must be stopped
and in LL0. NOTE: Value Not Saved is displayed if a value is
scrolled and ENTER/CLEAR is not pressed. Press NEXT to
continue programming.
Warning: If the pump is dropped or hit, inspect it for
damage. Do not use a pump that is damaged or is not
functioning properly.
Main Screen
Press NEXT.
Enter Reservoir Volume
1.
Press
or
to select desired Reservoir Volume.
2.
Press ENTER/CLEAR.
3.
Press NEXT
Enter Units
4.
Press
or
to select desired units.
5.
Press ENTER/CLEAR.
6.
Press NEXT.
Enter Concentration
Note: This screen does not appear if programming in
milliliters.
7.
Press
or
to select desired Concentration.
8.
Press ENTER/CLEAR.
9.
Press NEXT.
Enter Continuous Rate
10.
Press
or
to select desired Continuous Rate (select
the upper limit if the program will be adjusted in LL1).
11.
Press ENTER/CLEAR.
12.
Press NEXT.
Warning: This pump delivers medication at a constant rate
and/or allows delivery of a bolus dose at a specified time
interval. Programming the pump at a delivery rate other
than what is prescribed will cause over or under delivery of
medication.
Enter Demand Dose
13.
Press
or
to select desired Demand Dose (select the
upper limit if program will be adjusted in LL1).
14.
Press ENTER/CLEAR.
15.
Press NEXT.
*Enter Dose Lockout
Warning: When you enter a new value, any lockout time in
effect will be cleared. A demand dose could be requested
immediately upon starting the pump, resulting in over-
delivery.
16.
Press
or
to select desired Demand Dose Lockout.
Programming / Set-up sequence