17.
Press ENTER/CLEAR.
18.
Press NEXT.
Enter Doses Per Hour
Note: This screen will only appear if you have programmed
a demand dose and dose lockout is less than 1 hour.
19.
Press
or
to select desired Doses Per Hour.
20.
Press ENTER/CLEAR.
21.
Press NEXT.
*Clear Doses Given
22.
Press ENTER/CLEAR to clear the value for the number
of doses given. The display will show 0.
23.
Press NEXT.
*Clear Doses Attempted
24.
Press ENTER/CLEAR to clear the value for the number
of doses attempted by the patient.
The display will show 0.
25.
Press NEXT.
Clear Given (ml, mg, mcg)
26.
Press ENTER/CLEAR to clear the given value.
The display will show 0.00.
27.
Press NEXT.
Verify Air Detector Status
28.
Verify the setting is correct. (To change setting, see
Biomed Functions section.)
29.
Press NEXT.
*These screens will appear only if you have programmed
a demand dose.
Verify Upstream Sensor Status
30.
Verify the setting is correct. (To change setting, see
Biomed Functions section.)
31.
Press NEXT
Verify Programming
32.
Press NEXT repeatedly to review program.
To Operate in LL1 with Upper Limits, Decrease
Continuous Rate and/or Demand Dose
If pump will be operated in LL1 to allow adjustment of
Continuous Rate and/or Demand Dose (up to the maximum
entered in LL0):
33.
Change lock level to LL1.
34.
Press NEXT to go to Continuous Rate or Demand Dose
screen.
35.
Press
to select desired starting value.
36.
Press ENTER/CLEAR.
Programming / Set-up sequence - Continued
IN19882EN-112017.indd 3
11/01/18 11:49