Chapter 4: Printer and Computer Interface
4-2
Oximeter Operation and Service Manual
4. Using the PC adapter cable (part #3339), connect the 9-pin connector
of the PC adapter cable to the COM port on the PC as shown in figure
4-1.
2
WARNING! When connecting this monitor to any instru-
ment, verify proper operation before clinical use. Refer to
the instrument’s user manual for full instructions. Acces-
sory equipment connected to the monitor’s data interface
must be certified according to the respective IEC stan-
dards, i.e., IEC 950 for data processing equipment or IEC
601-1 for electromedical equipment. All combinations of
equipment must be in compliance with IEC 601-1-1 sys-
tems requirements. Anyone connecting additional equip-
ment to the signal input port or signal output port
configures a medical system, and, therefore, is responsible
that the system complies with the requirements of the sys-
tem standard IEC 601-1-1.
Figure 4-1: Connect the Oximeter to a PC
Connect to
Oximeter
Cable (catalog # 3350)
Cable (catalog # 3339)
Printer Adapter
db-9 Connector
PC Adapter
db-9 Connector
Connect to IBM
Compatible PC
or Printer