Chapter 9: Service Maintenance and Repair
Oximeter Operation and Service Manual
9-5
CON2,
CON3
Probe LED’s driving signals.
/DSP-
EN
Enable display drivers signal.
DSP-LD
Signal used to load data into display drivers.
FET-ON
Used to control MOSFET H-bridge Q8 and Q9, which powers
probe LED’s.
INTGAT
Signals used to enable integrator U27B.
IR-DRV
Used to control MOSFET H-bridge Q8 and Q9, which powers
probe LED’s.
/NMI
Non-Maskable Interrupt that signals watch dog timer time-out.
PON
Power on signal.
CLK
Clocks data into digital potentiometer chip U5.
POT-LD
POT-LD is used to select digital potentiometer chip U5.
SDI
Serial Data Input to digital potentiometer chip U5.
PRB-
DET
PRB-DET is used to inform processor if probe is plugged in.
/PRB-
FAULT
Signal used to inform processor about probe cable problem.
QUIET
Signal used to shut down power down supply during data acquisi-
tion.
RX
External RS-232 receive (CTS0) line.
RED-
DRV
Used to control MOSFET H-bridge Q8 and Q9, which powers
probe LED’s.
RESET
System reset signal generated by the watch dog timer U18.
RST-
INT
Resets integrator U27B by shorting capacitor C50.
SHT-
DWN
Output of the microprocessor to turn unit off.
TX
External RS-232 transmit line.
/TEN
Signal which enables RS-232 driver. Is connected to PRB-DET
signal.
Содержание 1818R
Страница 3: ......
Страница 7: ...This page is intentionally left blank ...
Страница 15: ...This page is intentionally left blank ...
Страница 29: ...This page is intentionally left blank ...
Страница 39: ...This page is intentionally left blank ...
Страница 55: ...This page is intentionally left blank ...
Страница 57: ...This page is intentionally left blank ...
Страница 58: ......