Chapter 4: Printer and Computer Interface
4-4
Oximeter Operation and Service Manual
Printing Spot Check Data
1. Set up the oximeter and the
printer as previously described.
2. Disconnect the SpO
2
sensor
from the oximeter.
3. Turn on the printer.
4. Turn on the oximeter. The
oximeter prints the spot check
data for each patient, from
patients 1-99, as shown in the
sample printout.
5. If there is no spot check data at
all, the message **** is printed.
6. The oximeter does not automati-
cally turn itself off when print-
ing spot check data.
7. If no valid data is collected for a
patient number, **** is printed.
8. If valid data is collected for a
patient number for less than one
minute, only the last measure-
ment is printed.
9. If data is collected for a patient number for more than one minute, the
relative time since the first measurement is shown for that patient.
10. Dashes indicate invalid or unavailable data (for example, the patient’s
finger was removed from the SpO
2
sensor).
Figure 4-2: Print Spot Check Data