Chapter 9: Service Maintenance and Repair
Oximeter Operation and Service Manual
9-7
1. Remove the three “C” cell batteries.
2. Connect the DC power supply’s “+” output to J9-1.
3. Connect the DC power supply’s “-” or “ground” output to J9-2.
Voltage Test Points
Unless otherwise noted, all voltages are measured with respect to ground at
P1-2.
CAUTION! Be sure to observe the polarity of the battery
connections. Failure to observe the polarity may cause fuse
F1 to blow.
Location
Condition
Nominal and Range
T19
Oximeter on.
5 VDC, ± 0.1 VDC
ANA+5 (T18)
Same as T19.
Same as T19.
FLTR+5
Same as TP14.
Same as T19.
ANA-5
Same as TP15.
Same as TP15.
FLTR-5 (T16)
Same as TP15.
-4.8 VDC to –6.2 VDC
V-REF (T17)
Oximeter on.
2.5 VDC
V-RAM
Oximeter on.
Oximeter off.
T19 minus 0.7 VDC.
V_BT minus 0.7 VDC.
U11-71
Oximeter on.
3.072 MHz, 50% duty
cycle.
Low battery LED.
When battery voltage
drops to 3.00 VDC.
Low battery LED first
turns on at 3.00 VDC,
±0.2 VDC.
Low battery shut off.
Oximeter on, battery
voltage slowly drop-
ping.
Oximeter shuts off
when battery voltage
drops to 2.7 VDC, ± 0.2
VDC.