Chapter 4: Printer and Computer Interface
Oximeter Operation and Service Manual
4-1
Chapter 4: Printer and Computer Interface
Computer Interface Description
The oximeter can transfer stored data to a PC. Information Data Log and
Spot Checks will be displayed/saved to a file in the format as shown in the
Spot Check Printouts
section of this manual.
Equipment Required
1. Hand Held Oximeter
2. 3350 Printer Cable
3. 3339 PC Adapter Cable
4. Communication settings:
Interface Instructions
1. Disconnect the sensor from the oximeter.
2. Connect the printer adapter cable (part #3350) to the sensor connector
on the unit.
3. Connect 9-pin connector of PC adapter cable (part #3350) to the
printer adapter cable (part #3339). Be sure to connect the 3339 to the
end of the 3350 cable marked as “printer.”
Baud rate
9600
Data bits
8
Stop bits
1
Parity
none
I/O Port:
Serial RS-232C
Data Type:
ASCII
Data Format:
9600 baud, 1 start bit, 8 data
bits, 1 stop bit, no parity
I/O Connector:
Standard DB-9