Chapter 9: Service Maintenance and Repair
Oximeter Operation and Service Manual
9-11
Remove all scope connections from the oximeter board.
Functional Testing
Disconnect the oximetry test fixture. Turn oximeter board off and then on
again. Connect Finger probe to oximeter board and put your finger in the
probe. Verify SpO
2
and Pulse Rate readings are displayed.
Verify that the pulse tone is heard when the pulse strength bar lights up.
Remove finger from finger probe. Verify “dashes.”
Measurement at T24
2V/DIV
0.4mS/Div
0V
0V