ALARM ID
DESCRIPTION
USER ACTION
AF:26
Water being taken into chamber at
wrong time. Cold water intake
valve “EVF” may have failed.
Check the water supply:
Check that the water supply pressure is within the permitted
range.
Follow the
DEFAULT PROCEDURE
described above. If the alarm
does not disappear, follow the
RESET PROCEDURE.
Also check that there are no leaks from the appliance
(no water
around it).
If there are leaks: turn off all water supply taps at once and
contact the After-Sales Service.
AF:28
Demineralised water being taken
into chamber at wrong time. Cold
water intake valve “EVF” may have
failed.
Check the water supply:
Check that the demineralised water supply pressure is within the
permitted range.
Follow the
DEFAULT PROCEDURE
described above. If the alarm
does not disappear, follow the
RESET PROCEDURE.
Also check that there are no leaks from the appliance
(no water
around it).
If there are leaks: turn off all water supply taps at once and
contact the After-Sales Service.
AF:29
Chamber does not empty. No
drainage.
Check the connection to the water drain,
especially that the height of the drain connections complies with
the specified requirements and that there are no restrictions in
the drain hoses.
AF:30
During the working cycle, the
chamber water level exceeds the
safety level. Water safety level.
Check the water supply to the appliance:
1. Intake pressure.
2. That the connections are correct, as specified in this manual.
For the user: follow the
DEFAULT PROCEDURE
described above.
If the alarm does not disappear, follow the
RESET PROCEDURE.
If there are leaks: turn off all water supply taps at once and
contact the After-Sales Service.
AF:32
Water standing in washing
chamber with appliance in
standby.
For the user: follow the
DEFAULT PROCEDURE
described above.
If the alarm does not disappear, follow the
RESET PROCEDURE.
If there are leaks: turn off all water supply taps at once and
contact the After-Sales Service.
AF:33
Steam condenser water intake
anomaly. No water in steam
condenser.
Indicates that there is no water in the steam condenser when
there should be.
Check the water supply:
1. Intake water pressure
2. That the connections are correct, as specified in this manual.
3. Blockages or restrictions in hoses.
Follow the
DEFAULT PROCEDURE
described above. If the alarm
does not disappear, follow the
RESET PROCEDURE.
AF:34
Steam condenser water drainage
failure.
Check that the drain connections are correct:
1. Height and position of the drain connection
2. That the connections are correct, as specified in this manual.
3. Blockages or restrictions in hoses.
For the user: follow the
DEFAULT PROCEDURE
described above.
If the alarm does not disappear, follow the
RESET PROCEDURE.