User Manual
GW2145 - GW1160 - GW4060 - GW4190
Page 43 - 86
Set
PARAMETER
DESCRIPTION
POSSIBLE USER ACTION
With set parameter
“-d=0”
With set
parameter
“-d=1”
-d
Superuser dosage setting disabling parameter.
(Parameter introduced from Main Firmware version
4.6.xx.yy).
If “-d: 0” the superuser is able to modify the dosages
If “-d: 1” the superuser is not able to modify the dosages
but can display them.
Default value: “-d: 0”.
This Parameter can only be modified by the authorised
engineer.
Display Only
Display Only
-E
Parameter for setting presence or absence of the WD-
LANE accessory.
(Parameter introduced from Main Firmware version
4.6.xx.yy).
Sets the serial port communication speed
“-E: Pr” = 9600 bps (for Printer connection)
“-E: Ln” = 11500 bps (for connection to WD-LANE - LAN
accessory).
Display and
Modification
Display and
Modification
-t
Test cycle
(Parameter Introduced from the Main Firmware version
1.11.0.33) The parameter can be used by authorized
technicians only; it allows the display and the start of a
Test program, on the display "
Pr t
". This test program
performs a short washing cycle and activates all the loads
of the device. The cycle is defined to be used for machine
verification after components replacement.
- Parameter value "--", you do not have access to the cycle
"
Pr t
"
- Parameter Value "on", the technician gets access to the
Test Program "Pr t".
In order not to run the Test cycle "
Pr t
": switch Off the
device to reset its condition.
Display Only
Display Only
In the following pictures some examples of "
Set
" parameters are featured.
fig. 31 – "
Set - C
" parameter. The user can fix the use of the powder dispenser (= "
dd
") or the peristaltic pump (= "
P1
"). When "
dd
" is chosen, a
possible value saved in "
Set - 1
" will be ignored by the system.