4.6
BRIEF TABLE OF THE PROGRAMS INSTALLED
For details of the various washing program phases, always refer to the enclosed
“
GW2145-GW1160-GW4060-
GW4190 Program Table
”
; the brief table below is purely guideline. The programs listed below may be subject to
modification for the purposes of improvement.
PROGRAMS with THERMAL DISINFECTION
In programs normally used in
GW
Glassware Washers,
thermal disinfection
takes place at the
beginning of the washing cycle
, for the most effective removal of contamination and grease from
the material for processing.
This process is different from the one used and recommended in Hospital Instrument Washers,
where this phase concludes the cycle so that the instruments are maintained in a disinfected state.
CUSTOM PROGRAMS
The
Custom programs
are programs which can be modified to meet specific requirements.
Programs can only be modified using the Smeg TRACELOG software; contact Smeg for
information concerning the use and installation of the program.
ID
GW - GLASSWARE WASHER
PR 1
Prewash
PR 2
Plastics 75°C [ + Drying ]
PR 3
Standard glassware 80°C [ + Drying ]
PR 4
Intensive glassware 90°C [ + Drying ]
PR 5
Agar Intensive glassware 93°C [ + Drying ]
PR 6
Thermal disinfection 93°C – 3 minutes [ + Drying ]
PR 7
Thermal disinfection 93°C – 10 minutes [ + Drying ]
PR 8
Thermal disinf intensive wash 93°C – 5 minutes [ + Drying ]
PR 9
Thermal disinf intensive wash 93°C – 10 minutes [ + Drying ]
PR 10
Custom 1 (default: Petroleum Grease, 93°C – 1 minute) [ + Drying ]
PR 11
Custom 2 (default: Mineral Oil, 93°C – 10 minutes [ + Drying ]
PR 12
Custom 3 (default: Petrol, 93°C – 5 minute) [ + Drying ]
PR 13
Custom 4 (default: Diesel, 93°C – 1 minute) [ + Drying ]
PR 14
Custom 5 (default: Intensive Petrol, 93°C – 10 minutes [ + Drying ]
PR 15
Custom 6 (default: Appliance auto-disinfection 93°C – 1 minute) [ + Drying ]
PR 16
Service - operation of peristaltic pumps, intake of det rinsing
Tab.
1
– Brief table
of the programs installed
DEFAULT DETERGENT DOSES AND WATER INTAKES
FOR SERVICE PROGRAM ONLY,
AD HOC DETERGENT DOSES AND WATER INTAKES
P1 = 5ml/litre – alkaline detergent
P1 = 12ml/litre
P2 = 3ml/litre - acid
P2 = 12ml/litre
P3 = 10ml/litre - degreasing
P3 = 12ml/litre
P2
a
= 4ml/litre (non-standard P2 doses used in special
cases).
P2
a
- Parameter not used in the Service program.
Water intake per phase, default = 9 litres
Water intake, phase 1 and phase 2 = 6 litres
Water intake, phase 3 = 9 litres
Tab.
2
– Brief table of doses and water intakes per phase.
MAINS WATER - DEMINERALISED WATER in PROGRAMS
The enclosed table indicates when a specific phase takes in cold water and when demineralised
water.
If demineralised water is deselected by means of the appliance setup parameters, the appliance
will always take in cold water.
Whenever possible, connection to demineralised water is recommended, for better final rinsing
results and to avoid the formation of limescale.