MENTOR 12 USER’S MANUAL
SECTION 4: CONFIGURATION – Page 93
Proceed as follows:
•
Press on the button that indicates the low level output you wish to use. This
will stay marked.
•
If you also wish to configure the presentation mode (end of scale and units)
that this output should be the equivalent to, press the button marked “…” of
each one and proceed as follows.
•
Select the unit in which you want to see your output, pressing “Volts” or
“Amps”.
•
Enter the ratio you wish to use for the parameter on the relative window. Bear
in mind that the end of scale of the Low Level Output is 7.07 V RMS. If you
select a ratio of 1000: 1 on the relative CONTROL display you would see
7070.000 V or A, according to the selected unit, when it is at maximum
regulation. This value is indicated below the ratio entry window.
•
Enter the correction value of the phase angle you wish to assign to the Low
Level Output on the relative window (necessary to simulate the error of a
converter or of a possible external amplifier).
•
Press OK to exit and confirm the changes or CANCEL to exit without
modifying anything.
Exactly the same process described is followed to configure the Low Level Outputs and
Sources in the Voltage section.
Once you have made the desired changes, press the button marked “DONE” if you wish
to preserve them or “CANCEL” if you wish to exit without executing the changes.
You will return to the General Screen where the new current configuration will be
displayed, if appropriate.