Chapter 3 Alarm Indication and Troubleshooting
HRS Series
3.3 Troubleshooting
3-113
AL36: DI sensor error
<Detection method> This alarm goes off when the value detected by the electric resistivity sensor is out
of -
0.6 to 5.6 MΩ range.
The value detected by the electric resistivity sensor is out of -
6.2 to 56.2 MΩ range.
- This alarm is set to be OFF as default.
- This alarm goes off when the electric resistivity sensor or the electric conductivity sensor
that is available as a separately sold accessory is used.
[Part numbers of the separately sold accessories]
HRS-DI001, HRS-DI003, HRS-DI004, HRS-DI005, HRS-DI008, HRS-DI009
- Refer to the Operation Manual that is attached to the separately sold accessory.
Is the electric
resistivity sensor or electric
conductivity sensor connected
to the product?
Connect the sensor.
Replace the electric resistivity
sensor or electric conductivity
sensor.
N
Y
DI sensor error
AL36