3.2.5 Alert indicator light

The CMJ-01 interface unit features a “Visual Alert” light on the front panel (see drawing 3).

3.2.6 Master Reset

For single TL-2015R2, to reset its alert, simply pressing on the “Acknowledge” button or using 

Caregiver key

  

(CK-01) (Refer to section 2.1.1), or using a wireless reset button which is programmed to “Master  
Reset” function.

For multiple TL-2015R2 connected together, in order to reset all TL-2015R2, one should use “Master Reset” 
button to reset all of them at the same time, otherwise reset each by pressing the “Acknowledge” button or using 

Caregiver key

 (Refer to section 2.1.1).

On drawing 9, each TL-2015R2 is located at different zone with a master reset button nearby. Pressing on the 
master reset button will reset the same zone’s TL-2015R2, and then the TL-2015R2 transfers it to the  
next TL-2015R2. 

3.2.7 Frequencies of Transmission and Reception

For TL-2015R2, the frequencies of transmission (TX) and reception (RX) should be different, or its transmission 
may be interference by other transmitters. This is also applied to the extension’s receivers through CMJ-01. 
(Please refer to the bottom of TL-2015R2 for the frequency coniguration of the device.)

4.0 APPROVALS

For information, Smart digital wireless transmitters and receivers are FCC certiied and comply with 
Telecommunication (Low Power Devices) standards throughout the world. Our products are also CE registered 
and ETL (UL equivalent). This product is manufactured to ISO9001 (year 2000) Standard.  

5.0 SUMMARY OF FEATURES AND FUNCTIONS:

5.1  

The TL-2015R2 unit has a large, 2-line LCD display

5.2  

Selectable automatic out of range (OORC) and interference check with indication.

5.3  

Automatic AC main power supply failure detection with indication

5.4  

Indicates and can send on automatic fault alerts from other Smart equipment 

5.5  

Standard pre-prepared (plug-in) interconnection cables

5.6  

Compatible with other Smart systems and monitoring equipment

5.7  

Data outputs for system integration e.g. PC, printer, etc

5.8  

Selectable 

Caregiver key

 , or manual button Reset function for local operation

5.9  

Programming port

5.10   Adjustable volume

Содержание TL-2015R2

Страница 1: ...nd Nurse Call Paging System InstallationandUseInstructions 433 PGD Caegiver Pocket Pager with LCD Display 433 NC Wireless Nurse Call Button Optional Accessories SMART CAREGIVER C O RPO RAT I O N Tel 8...

Страница 2: ...2 Large Facility Central Monitor TL 2015R2 Connection Diagrams for Large Facility Central Monitor TL 2015R2...

Страница 3: ...SETTING UP AND INSTALLING THE CARE STATION 3 0 OPTIONAL INTERFACE EXPANSION UNIT 4 0 APPROVALS 5 0 SUMMARY OF FEATURES AND FUNCTIONS 6 0 TECHNICAL SPECIFICATIONS 7 0 USER INSTRUCTIONS 8 0 LIST OF COMP...

Страница 4: ...om ield equipment by providing different sound and display indications In addition a stand alone wireless transponder repeater unit WBUM 02 is available that can substantially increase the distance at...

Страница 5: ...ot work properly The hand held tester unit may help locate the source in this case c Testing the wireless transmission on the site Up to 80 metre 250ft wireless range between devices is normal for Sma...

Страница 6: ...op bit and no parity bit And the output looks like this Toilet Alert ID 010123 2011 01 26 16 27 06 2 1 6 Data In Out This is a spare port for future development 2 1 7 Relay O P May be used for Normall...

Страница 7: ...res The length of such cable extension can be up to 250 metres Each ield receiver can detect data signals from an area having a radius up to 50 metres 150ft subject to survey and system range testing...

Страница 8: ...terface interconnection cables If the interface unit CMJ 01 is to be employed and only it s incorporated ield data receiver is to be utilised a standard pre prepared cable connects between the Interfa...

Страница 9: ...ifferent or its transmission may be interference by other transmitters This is also applied to the extension s receivers through CMJ 01 Please refer to the bottom of TL 2015R2 for the frequency conigu...

Страница 10: ...ximum rating for relay output contacts 1 amp e g for door lock 6 Display Simultaneous alphanumeric text for several detected patient transmitter units These alerts rotate in the display window 7 User...

Страница 11: ...central monitor OORC system auto checking function if enabled Used to detect if a ield unit becomes faulty or wireless interference is sensed This is reported and indicated at the TL 2015R2 7 3 1 Wire...

Страница 12: ...cordless PAD you must irst is to register its ID in to TL 2015R2 So it needs the following functions to operate the registration Enter into setup mode You must use the caregiver key for this Press the...

Страница 13: ...S N number ID of TM 04 and 00 is the irst number room assigned for this ID Use A Key to increate the room number its range is from 00 to 25 Once conirm the number use C Key to enter to next display Sa...

Страница 14: ...its range is from 0 to 9 XXXX serial number ID of the recorded TM 04 Use A Key to show next record When it show QUIT applying Caregiver key can then quit the list function Or when it is showing a emp...

Страница 15: ...n added to the working list memory The TL 2015R2 is now checking this pad continuously looking for the handshake signal every 30 seconds If it does not see this signal for 10 minutes it will show the...

Страница 16: ...l Time Clock setup Real time clock setup function is under the SETUP menu use section 7 4 to go into setup mode And then select Clock Time 7 5 1 Change Date Use Caregiver key entering into the clock s...

Страница 17: ...using for TL 2015R2 it shows 203 only that TL 2015R2 interprets the room number as 20 and the bed number is 3 So in order to make it display correctly in TL 2015R2 for BTX 02 programming that at the R...

Страница 18: ...let button 007 Toilet button at position 007 LCD Display Toilet Alarm Area No 007 4 Wristband transmitter s Calls can be reset at the central monitor Note Due to the mobility of these units there is n...

Страница 19: ...number LCD Display Door Fault 02 Other faults that are indicated a Power loss on units with batteries Bed mon AC loss 203 power is lost on this unit Only sent once since can be used battery only LCD...

Страница 20: ...3004SYS TL 3005SYS Door monitor with LCD display and external strip receiver b TL 3004SYS TL 3005SYS Easy installed door strip monitor with no display display available on Central Monitor c TL 2012SR...

Страница 21: ...21 Inter connections to TL 2015R2 TL 2015R2 TL 2015R2...

Страница 22: ...22 TL 2015R2 TL 2015R2 Back View...

Страница 23: ...23 TL 2015R2 TL 2015R2 Interface Unit...

Страница 24: ...24 TL 2015R2 TL 3005 TL 2012SREV...

Страница 25: ...25 TL 2015R2 TL 2015R2 Back View TL 3005 TL 2012SREV TL 3005...

Страница 26: ...26 TL 2015R2 TL 3005 TL 2012SREV...

Страница 27: ...27 TL 2015R2 1 TL 2015R2 2 TL 2015R2 3 Multi TL 2015R2 Application...

Страница 28: ...28 TL 2015R2 TL 2015R2...

Страница 29: ...29 TL 2015R2 3 Multi TL 2015R2 Application TL 2015R2 1 TL 2015R2 2...

Отзывы: