7.2 Acceptance and reset of an alert

When the TL-2015R2 receives an alert signal, it is automatically displayed in the LCD showing type of alert with 
ield unit identity number. A pulsing sounder alerts the user that there is an alert.

To reset an active alert at the TL-2015R2, press the manual reset button on the front panel and then attend to the 
patient calling. 

When the reset button is pressed the system will rotate any alerts on the display for approximately 10 seconds 
before the display is cleared as a reminder of alerts on the system.

A bed monitor return signal will remain on the display to remind you that a patient is moving and will issue  
two beeps.

Some alert units such as a bed head alert, door monitor and bed monitor, will remain in an alert state until reset 
at the source so resetting at the TL-2015R2 central monitor will only be temporary and you must reset at the unit 
to clear the alert.

In the case where it is desired to have greater security on the TL-2015R2, the reset button may be disabled 
during installation. In this case only the 

Caregiver key

 can be used by the caregiver to reset the alert, and the 

reset button is disabled. 

7.3 The central monitor OORC system auto checking function (if enabled)

Used to detect if a ield unit becomes faulty, or wireless interference is sensed. This is reported and indicated at 
the TL-2015R2.

7.3.1 Wireless interference detected

The TL-2015R2 will display “Interference Detected” and emit a “beep-beep” sound. This will continue until the 
interference is removed.

7.3.2 Field unit fault detected

Field units automatically send a “report” signal to the TL-2015R2 approximately every 6 hours. 
With the OORC switch on, if a ield unit does not report by transmitting 4 consecutive signals (in a period of 24 
hours) the central monitor will display (for example) “Bed Head Fault” together with the particular ield unit 4  
digit identity number.  This will be displayed until the TL-2015R is reset (see 2.1.1 above).

If this ield unit continues to be faulty, or “out of range” (no report signal), the central monitor will again display 
and indicate a fault after a further 6 hours.

In the case of a ield unit being removed from the system, it needs to be removed from the TL-2015R2 memory. 
This is simply done by turning “off” the OORC switch, and then switching back “on” (if continued auto fault 
checking required).

With the OORC switch turned “on”, any new ield unit/identity being added to the system will automatically be 
added to the central monitor fault checking memory with its irst alert transmission.  

Содержание TL-2015R2

Страница 1: ...nd Nurse Call Paging System InstallationandUseInstructions 433 PGD Caegiver Pocket Pager with LCD Display 433 NC Wireless Nurse Call Button Optional Accessories SMART CAREGIVER C O RPO RAT I O N Tel 8...

Страница 2: ...2 Large Facility Central Monitor TL 2015R2 Connection Diagrams for Large Facility Central Monitor TL 2015R2...

Страница 3: ...SETTING UP AND INSTALLING THE CARE STATION 3 0 OPTIONAL INTERFACE EXPANSION UNIT 4 0 APPROVALS 5 0 SUMMARY OF FEATURES AND FUNCTIONS 6 0 TECHNICAL SPECIFICATIONS 7 0 USER INSTRUCTIONS 8 0 LIST OF COMP...

Страница 4: ...om ield equipment by providing different sound and display indications In addition a stand alone wireless transponder repeater unit WBUM 02 is available that can substantially increase the distance at...

Страница 5: ...ot work properly The hand held tester unit may help locate the source in this case c Testing the wireless transmission on the site Up to 80 metre 250ft wireless range between devices is normal for Sma...

Страница 6: ...op bit and no parity bit And the output looks like this Toilet Alert ID 010123 2011 01 26 16 27 06 2 1 6 Data In Out This is a spare port for future development 2 1 7 Relay O P May be used for Normall...

Страница 7: ...res The length of such cable extension can be up to 250 metres Each ield receiver can detect data signals from an area having a radius up to 50 metres 150ft subject to survey and system range testing...

Страница 8: ...terface interconnection cables If the interface unit CMJ 01 is to be employed and only it s incorporated ield data receiver is to be utilised a standard pre prepared cable connects between the Interfa...

Страница 9: ...ifferent or its transmission may be interference by other transmitters This is also applied to the extension s receivers through CMJ 01 Please refer to the bottom of TL 2015R2 for the frequency conigu...

Страница 10: ...ximum rating for relay output contacts 1 amp e g for door lock 6 Display Simultaneous alphanumeric text for several detected patient transmitter units These alerts rotate in the display window 7 User...

Страница 11: ...central monitor OORC system auto checking function if enabled Used to detect if a ield unit becomes faulty or wireless interference is sensed This is reported and indicated at the TL 2015R2 7 3 1 Wire...

Страница 12: ...cordless PAD you must irst is to register its ID in to TL 2015R2 So it needs the following functions to operate the registration Enter into setup mode You must use the caregiver key for this Press the...

Страница 13: ...S N number ID of TM 04 and 00 is the irst number room assigned for this ID Use A Key to increate the room number its range is from 00 to 25 Once conirm the number use C Key to enter to next display Sa...

Страница 14: ...its range is from 0 to 9 XXXX serial number ID of the recorded TM 04 Use A Key to show next record When it show QUIT applying Caregiver key can then quit the list function Or when it is showing a emp...

Страница 15: ...n added to the working list memory The TL 2015R2 is now checking this pad continuously looking for the handshake signal every 30 seconds If it does not see this signal for 10 minutes it will show the...

Страница 16: ...l Time Clock setup Real time clock setup function is under the SETUP menu use section 7 4 to go into setup mode And then select Clock Time 7 5 1 Change Date Use Caregiver key entering into the clock s...

Страница 17: ...using for TL 2015R2 it shows 203 only that TL 2015R2 interprets the room number as 20 and the bed number is 3 So in order to make it display correctly in TL 2015R2 for BTX 02 programming that at the R...

Страница 18: ...let button 007 Toilet button at position 007 LCD Display Toilet Alarm Area No 007 4 Wristband transmitter s Calls can be reset at the central monitor Note Due to the mobility of these units there is n...

Страница 19: ...number LCD Display Door Fault 02 Other faults that are indicated a Power loss on units with batteries Bed mon AC loss 203 power is lost on this unit Only sent once since can be used battery only LCD...

Страница 20: ...3004SYS TL 3005SYS Door monitor with LCD display and external strip receiver b TL 3004SYS TL 3005SYS Easy installed door strip monitor with no display display available on Central Monitor c TL 2012SR...

Страница 21: ...21 Inter connections to TL 2015R2 TL 2015R2 TL 2015R2...

Страница 22: ...22 TL 2015R2 TL 2015R2 Back View...

Страница 23: ...23 TL 2015R2 TL 2015R2 Interface Unit...

Страница 24: ...24 TL 2015R2 TL 3005 TL 2012SREV...

Страница 25: ...25 TL 2015R2 TL 2015R2 Back View TL 3005 TL 2012SREV TL 3005...

Страница 26: ...26 TL 2015R2 TL 3005 TL 2012SREV...

Страница 27: ...27 TL 2015R2 1 TL 2015R2 2 TL 2015R2 3 Multi TL 2015R2 Application...

Страница 28: ...28 TL 2015R2 TL 2015R2...

Страница 29: ...29 TL 2015R2 3 Multi TL 2015R2 Application TL 2015R2 1 TL 2015R2 2...

Отзывы: