Waveform: Symmetrical bi-phasic square waveform when measured into a
resistive load.
Description of your controller’s symbols:
There are a number of technical markings on your controller. These can be
explained as follows:
The controller and belt are manufactured for Bio-Medical Research Ltd,
Parkmore Business Park West, Galway, Ireland.
The controller requires 1 x 3.6 volt (NiMH) DC battery pack DC is indicated by
the symbol:
The Output Frequency indicates the number of pulses per second transmitted
by the controller. This is measured in hertz, which is indicated by “Hz”.
Output (RMSA) means the maximum output root mean square current for each
channel.
Output (RMSV) means the maximum output root mean square voltage for each
channel.
This symbol means “Attention, consult the accompanying documents”.
This symbol means type BF applied parts (electrodes).
This symbol on your product is to indicate conformity to the requirements
of the Medical Device Directive (93/42/EEC). 0086 is the number of
the notified body (BSI). The CE mark applied also indicates that this
equipment complies with the requirements of the ROHS Directive
(2011/65/EU) for the Restriction of Hazardous Substances in Electrical
and Electronic Equipment.
Conforms to UL STD 60601-1.
Certified to CAN/CSA.
STD C22.2 NO 601.1
SN stands for ‘serial number’. On the label on the back of the controller is the
serial number specific to this controller. The letter preceding the number indicates
the year of manufacture, where “T” denotes 2014, “U” denotes 2015 etc.
The belt’s batch number is represented on the belt packaging, by the number
corresponding with the LOT symbol.
Rated Outputs – Voltage/Currents: E:70 and X:70
Attention:
This product can deliver currents in excess of 2mA/cm
² if used with incorrect
gel pads.
At the end of the product lifecycle, do not throw this product or batteries
into the normal household garbage, but bring it to a collection point for
the recycling of electronic equipment.
Some product materials can be re-used if you bring them to a recycling
point. By re-using some parts or raw materials from used products you
make an important contribution to the protection of the environment.
Please contact your local authorities if you need more information about
collection points in your area.
Waste Electrical and Electronic Equipment can have potentially harmful
effects on the environment. Incorrect disposal can cause harmful toxins
to build up in the air, water and soil and can be harmful to human health.
Parameter
Output RMSV
Output RMSA
Output Frequency
DC Component: Approx.
Positive Pulse Width:
Negative Pulse Width:
Interphase Interval:
7.35V
14.7mA
50-70Hz
0 C
200-315µs
200-315µs
100µs
13.1V
13.1mA
50-70Hz
0 C
200-315µs
200-315µs
100µs
15.5V
10.3mA
50-70Hz
0 C
200-315µs
200-315µs
100µs
500
Ω
1K
Ω
1K5
Ω
19
English