6 Maintenance and inspection
Sirona Dental Systems GmbH
6.3 Replace the sensor cable
Operating Instructions and Installation XIOS XG USB module and sensors
Attach a replacement cable
1. Place the plug for the sensor cable properly on the sensor. Both parts
must interlock.
2. Screw the sensor cable to the sensor using the screws supplied.
Initially only tighten the screws until a light resistance can be felt. After
this tighten both screws carefully.
3. Cover the screw heads with a new protective screw cap. Press the
cap into the plug of the sensor cable until it locks in place.
The cable replacement is now completed. The sensor can be used
again.