4 Installation
Sirona Dental Systems GmbH
4.1 Fitting a second protective ground conductor on the PC
Operating Instructions and Installation XIOS XG USB module and sensors
4
Installation
4.1
Fitting a second protective ground conductor on
the PC
Depending on whether the PC is operated inside or outside of the patient
environment (up to 1.5 m around the patient), an additional protective
ground conductor must be provided on the PC housing.
Power transmission to the X-ray system
Definition of the patient environment in accordance with IEC 60601-1
Within the patient environment (A), direct contact is only permissible with
devices or system parts that are approved for use in the patient
CAUTION
Leakage currents from the PC are transmitted to the X-ray system.
If a PC is not sufficiently grounded, there is a risk of electric shock for
both the patient and the user.
➢ The PC must be connected to a grounded electric outlet when in
operation.
➢ If the PC is operated in the patient environment (up to 1.5 m around
the patient), it must also be fitted with an additional protective
ground conductor.