64 09 713 D3610
D3610
.
201.01.08
.
02 09.2016
19
Sirona Dental Systems GmbH
3 System description
Operating Instructions and Installation XIOS XG USB module and sensors
3.6 Intraoral X-ray tube assembly
båÖäáëÜ
3.6
Intraoral X-ray tube assembly
The following requirements apply to the intraoral X-ray tube assembly:
Intraoral X-ray tube assembly with multipulse technology (direct current)
0.14 – 1.4 mAs at 60 – 70 kV and 8" cone
This information must be modified accordingly for other tube lengths or
single-pulse X-ray tube assemblies.
For optimal image quality we recommend using a multipulse X-ray tube
assembly with a 12"cone.
IMPORTANT
The intraoral X-ray tube assembly must be installed in accordance with
the manufacturer's instructions and requirements. Please follow the
manual for your X-ray tube assembly.